Download
s12891-021-04933-y.pdf 1,38MB
WeightNameValue
1000 Titel
  • Is cement-augmented sacroiliac screw fixation with partially threaded screws superior to that with fully threaded screws concerning compression and pull-out force in fragility fractures of the sacrum? – a biomechanical analysis
1000 Autor/in
  1. Hack, Juliana |
  2. Safi, Maiwand |
  3. Bäumlein, Martin |
  4. Lenz, Julia |
  5. Bliemel, Christopher |
  6. Ruchholtz, Steffen |
  7. Oberkircher, Ludwig |
1000 Erscheinungsjahr 2021
1000 Publikationstyp
  1. Artikel |
1000 Online veröffentlicht
  • 2021-12-10
1000 Erschienen in
1000 Quellenangabe
  • 22(1):1034
1000 Copyrightjahr
  • 2021
1000 Lizenz
1000 Verlagsversion
  • https://doi.org/10.1186/s12891-021-04933-y |
  • https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8665623/ |
1000 Publikationsstatus
1000 Sprache der Publikation
1000 Abstract/Summary
  • Background!#!Providing a stable osteosynthesis in fragility fractures of the pelvis can be challenging. Cement augmentation increases screw fixation in osteoporotic bone. Generating interfragmentary compression by using a lag screw also improves the stability. However, it is not known if interfragmentary compression can be achieved in osteoporotic sacral bone by cement augmentation of lag screws. The purpose of this study was to compare cement-augmented sacroiliac screw osteosynthesis using partially versus fully threaded screws in osteoporotic hemipelvises concerning compression of fracture gap and pull-out force.!##!Methods!#!Nine fresh-frozen human cadaveric pelvises with osteoporosis were used. In all specimens, one side was treated with an augmented fully threaded screw (group A), and the other side with an augmented partially threaded screw (group B) after generating a vertical osteotomy on both sides of each sacrum. Afterwards, first a compression test with fracture gap measurement after tightening of the screws was performed, followed by an axial pull-out test measuring the maximum pull-out force of the screws.!##!Results!#!The fracture gap was significantly wider in group A (mean: 1.90 mm; SD: 1.64) than in group B (mean: 0.91 mm; SD: 1.03; p = 0.028). Pull-out force was higher in group A (mean: 1696 N; SD: 1452) than in group B (mean: 1616 N; SD: 824), but this difference was not statistically significant (p = 0.767).!##!Conclusions!#!Cement augmentation of partially threaded screws in sacroiliac screw fixation allows narrowing of the fracture gap even in osteoporotic bone, while resistance against pull-out force is not significantly lower in partially threaded screws compared to fully threaded screws.
1000 Sacherschließung
lokal Sacrum/surgery [MeSH]
lokal Partially threaded screw
lokal Fully threaded screw
lokal Cement augmentation
lokal Humans [MeSH]
lokal Bone Cements [MeSH]
lokal Biomechanical Phenomena [MeSH]
lokal Fragility fracture of the pelvis
lokal Osteoporosis
lokal Bone Screws [MeSH]
lokal Sacrum/diagnostic imaging [MeSH]
lokal Pelvis [MeSH]
lokal Research
1000 Liste der Beteiligten
  1. https://frl.publisso.de/adhoc/uri/SGFjaywgSnVsaWFuYQ==|https://frl.publisso.de/adhoc/uri/U2FmaSwgTWFpd2FuZA==|https://frl.publisso.de/adhoc/uri/QsOkdW1sZWluLCBNYXJ0aW4=|https://frl.publisso.de/adhoc/uri/TGVueiwgSnVsaWE=|https://frl.publisso.de/adhoc/uri/QmxpZW1lbCwgQ2hyaXN0b3BoZXI=|https://frl.publisso.de/adhoc/uri/UnVjaGhvbHR6LCBTdGVmZmVu|https://frl.publisso.de/adhoc/uri/T2JlcmtpcmNoZXIsIEx1ZHdpZw==
1000 Hinweis
  • DeepGreen-ID: 7ad7ff1e0a2a47219c800eb19fdfebd4 ; metadata provieded by: DeepGreen (https://www.oa-deepgreen.de/api/v1/), LIVIVO search scope life sciences (http://z3950.zbmed.de:6210/livivo), Crossref Unified Resource API (https://api.crossref.org/swagger-ui/index.html), to.science.api (https://frl.publisso.de/), ZDB JSON-API (beta) (https://zeitschriftendatenbank.de/api/), lobid - Dateninfrastruktur für Bibliotheken (https://lobid.org/resources/search)
1000 Label
1000 Dateien
1000 Objektart article
1000 Beschrieben durch
1000 @id frl:6443450.rdf
1000 Erstellt am 2023-04-26T18:52:47.320+0200
1000 Erstellt von 322
1000 beschreibt frl:6443450
1000 Zuletzt bearbeitet 2023-10-19T14:46:19.798+0200
1000 Objekt bearb. Thu Oct 19 14:46:19 CEST 2023
1000 Vgl. frl:6443450
1000 Oai Id
  1. oai:frl.publisso.de:frl:6443450 |
1000 Sichtbarkeit Metadaten public
1000 Sichtbarkeit Daten public
1000 Gegenstand von

View source