Download
s10439-021-02849-1.pdf 759,90KB
WeightNameValue
1000 Titel
  • Validation of a Miniaturized Test Loop for the Assessment of Human Blood Damage by Continuous-Flow Left-Ventricular Assist Devices
1000 Autor/in
  1. Woelke, Eva |
  2. Mager, Ilona |
  3. Schmitz-Rode, Thomas |
  4. Steinseifer, Ulrich |
  5. Clauser, Johanna |
1000 Erscheinungsjahr 2021
1000 Publikationstyp
  1. Artikel |
1000 Online veröffentlicht
  • 2021-08-24
1000 Erschienen in
1000 Quellenangabe
  • 49(12):3165-3175
1000 Copyrightjahr
  • 2021
1000 Lizenz
1000 Verlagsversion
  • https://doi.org/10.1007/s10439-021-02849-1 |
  • https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8671281/ |
1000 Publikationsstatus
1000 Sprache der Publikation
1000 Abstract/Summary
  • Despite improved hemocompatibility of left-ventricular assist devices (LVADs), assessment of blood damage remains mandatory in preclinical testing as standardized by ASTM-F1841. The most relevant test fluid is fresh, non-pooled human blood, but the limited volume of a standard donation requires significantly smaller loops than those commonly used with animal blood. In a recent study with porcine blood, we verified a miniaturized test loop with only 160 mL for the ASTM-conform paired testing of at least two LVADs and a static reference. Here, we validated this mini test loop for standardized assessment of blood damage with one 450-mL single donation of fresh human blood. Blood damage was assessed for HeartMate 3 and BPX-80 in 9 experiments with heparinized human blood for 6 hours. We analyzed plasma free hemoglobin, von Willebrand factor (vWF) concentration and collagen-binding functionality and calculated indices of hemolysis and vWF-ratios. Overall, we observed less blood damage compared to our previous study; however, the differences in mean indices of hemolysis and in mean normalized vWF-ratio between BPX-80 and HeartMate 3 were consistent for human blood. Thus, our mini test loop proved to be valid for preclinical standardized assessment of blood damage with only 450 mL of fresh human blood.
1000 Sacherschließung
lokal Hemolysis
lokal Swine [MeSH]
lokal Miniaturization [MeSH]
lokal Humans [MeSH]
lokal Hemocompatibility
lokal Animals [MeSH]
lokal Heart-Assist Devices/adverse effects [MeSH]
lokal Original Article
lokal Hemolysis [MeSH]
lokal von Willebrand factor
lokal Materials Testing [MeSH]
lokal Heartmate 3
lokal LVAD
1000 Liste der Beteiligten
  1. https://frl.publisso.de/adhoc/uri/V29lbGtlLCBFdmE=|https://frl.publisso.de/adhoc/uri/TWFnZXIsIElsb25h|https://frl.publisso.de/adhoc/uri/U2NobWl0ei1Sb2RlLCBUaG9tYXM=|https://frl.publisso.de/adhoc/uri/U3RlaW5zZWlmZXIsIFVscmljaA==|https://orcid.org/0000-0002-9050-2391
1000 Hinweis
  • DeepGreen-ID: c66cf4e200fa446dbc81bddb9a1915bb ; metadata provieded by: DeepGreen (https://www.oa-deepgreen.de/api/v1/), LIVIVO search scope life sciences (http://z3950.zbmed.de:6210/livivo), Crossref Unified Resource API (https://api.crossref.org/swagger-ui/index.html), to.science.api (https://frl.publisso.de/), ZDB JSON-API (beta) (https://zeitschriftendatenbank.de/api/), lobid - Dateninfrastruktur für Bibliotheken (https://lobid.org/resources/search)
1000 Label
1000 Dateien
1000 Objektart article
1000 Beschrieben durch
1000 @id frl:6446181.rdf
1000 Erstellt am 2023-04-28T12:25:46.448+0200
1000 Erstellt von 322
1000 beschreibt frl:6446181
1000 Zuletzt bearbeitet Fri Oct 20 17:44:24 CEST 2023
1000 Objekt bearb. Fri Oct 20 17:44:24 CEST 2023
1000 Vgl. frl:6446181
1000 Oai Id
  1. oai:frl.publisso.de:frl:6446181 |
1000 Sichtbarkeit Metadaten public
1000 Sichtbarkeit Daten public
1000 Gegenstand von

View source