Download
s00402-021-04036-2.pdf 799,54KB
WeightNameValue
1000 Titel
  • Imageless robotic handpiece-assisted total knee arthroplasty: a learning curve analysis of surgical time and alignment accuracy
1000 Autor/in
  1. Savov, Peter |
  2. Tuecking, Lars-Rene |
  3. Windhagen, Henning |
  4. Ehmig, Jonathan |
  5. Ettinger, Max |
1000 Erscheinungsjahr 2021
1000 Publikationstyp
  1. Artikel |
1000 Online veröffentlicht
  • 2021-07-14
1000 Erschienen in
1000 Quellenangabe
  • 141(12):2119-2128
1000 Copyrightjahr
  • 2021
1000 Lizenz
1000 Verlagsversion
  • https://doi.org/10.1007/s00402-021-04036-2 |
  • https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8595234/ |
1000 Publikationsstatus
1000 Sprache der Publikation
1000 Abstract/Summary
  • Introduction!#!Robotic-assisted surgery techniques are increasing in total knee arthroplasty (TKA). One crucial point is the prolonged time of surgery. The primary objective of this study was to determine the learning curve necessary to minimize the time of surgery. The secondary objective was to evaluate the accuracy of the implant alignment when using an imageless robotic system for TKA.!##!Materials and methods!#!In a case-control study, the first 70 consecutive robotic-assisted TKA procedures performed by a single senior surgeon were analyzed with regard to surgery time and implant alignment by comparing the intraoperative plan with the postoperative alignment. The evaluation of the learning curve with respect to surgery time was conducted using cumulative summation (CUSUM) analysis. The joint line height was measured with a new technique. Surgery time and joint line reconstruction were compared to 70 consecutive conventional TKA procedures.!##!Results!#!The learning curve for robotic TKA was completed after 11 cases. The learning curve did not influence the accuracy of joint line obliquity, joint line height, or limb alignment. The intraoperative plan designed for the robotic system was precisely implemented. The mean skin-to-skin time in the robotic group after the learning curve was completed did not differ from that in the manual group. A significant positive correlation was observed between the preoperative hip-knee-ankle angle and the postoperative distalization of the joint line in the robotic-assisted TKA group.!##!Conclusion!#!After completing the initial learning curve of 11 cases, the surgery time required to perform imageless robotic handpiece-assisted TKA was similar to that for the conventional technique. However, no learning curve was observed for the implant positioning when using the imageless robotic system. The implementation of the intraoperative plan was accurate up to < 2°. The precision of the system allows the implementation of different joint balancing approaches between valgus and varus morphotypes.
1000 Sacherschließung
lokal Knee Arthroplasty
lokal Robotic handpiece
lokal Humans [MeSH]
lokal Total knee arthroplasty (TKA)
lokal Computer-assisted surgery (CAS)
lokal Knee Joint/surgery [MeSH]
lokal Learning curve
lokal Arthroplasty, Replacement, Knee [MeSH]
lokal Learning Curve [MeSH]
lokal Robotic Surgical Procedures [MeSH]
lokal Case-Control Studies [MeSH]
lokal Kinematic alignment
lokal Operative Time [MeSH]
lokal New Technologies in Knee Arthroplasty
1000 Liste der Beteiligten
  1. https://orcid.org/0000-0003-4791-6051|https://frl.publisso.de/adhoc/uri/VHVlY2tpbmcsIExhcnMtUmVuZQ==|https://frl.publisso.de/adhoc/uri/V2luZGhhZ2VuLCBIZW5uaW5n|https://frl.publisso.de/adhoc/uri/RWhtaWcsIEpvbmF0aGFu|https://frl.publisso.de/adhoc/uri/RXR0aW5nZXIsIE1heA==
1000 Hinweis
  • DeepGreen-ID: f4194f133c9c4f76a97d791a265985b0 ; metadata provieded by: DeepGreen (https://www.oa-deepgreen.de/api/v1/), LIVIVO search scope life sciences (http://z3950.zbmed.de:6210/livivo), Crossref Unified Resource API (https://api.crossref.org/swagger-ui/index.html), to.science.api (https://frl.publisso.de/), ZDB JSON-API (beta) (https://zeitschriftendatenbank.de/api/), lobid - Dateninfrastruktur für Bibliotheken (https://lobid.org/resources/search)
1000 Label
1000 Dateien
1000 Objektart article
1000 Beschrieben durch
1000 @id frl:6447344.rdf
1000 Erstellt am 2023-04-28T15:47:25.419+0200
1000 Erstellt von 322
1000 beschreibt frl:6447344
1000 Zuletzt bearbeitet Fri Oct 20 20:29:32 CEST 2023
1000 Objekt bearb. Fri Oct 20 20:29:32 CEST 2023
1000 Vgl. frl:6447344
1000 Oai Id
  1. oai:frl.publisso.de:frl:6447344 |
1000 Sichtbarkeit Metadaten public
1000 Sichtbarkeit Daten public
1000 Gegenstand von

View source