Download
s00068-021-01805-x.pdf 1,10MB
WeightNameValue
1000 Titel
  • The effect of an infra-acetabular screw for anatomically shaped three-dimensional plate or standard plate designs in acetabulum fractures: a biomechanical analysis
1000 Autor/in
  1. Graul, I. |
  2. Marintschev, I. |
  3. Pizanis, A. |
  4. Herath, S. C. |
  5. Pohlemann, T. |
  6. Fritz, Tobias |
1000 Erscheinungsjahr 2021
1000 Publikationstyp
  1. Artikel |
1000 Online veröffentlicht
  • 2021-10-07
1000 Erschienen in
1000 Quellenangabe
  • 48(5):3757-3764
1000 Copyrightjahr
  • 2021
1000 Lizenz
1000 Verlagsversion
  • https://doi.org/10.1007/s00068-021-01805-x |
  • https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9532306/ |
1000 Publikationsstatus
1000 Sprache der Publikation
1000 Abstract/Summary
  • Background!#!Various plate shapes and implant configurations are used for stabilization of acetabulum fractures via anterior approaches. Little is known about the biomechanical stability of a two-dimensionally shaped 'conventional' plate ('J-Plate'-JP) in comparison to three-dimensionally shaped plate configurations (3DP). In addition, the augmentary effect of an infra-acetabular lag-screw (IACS) fixation for anterior column and posterior hemi-transverse acetabulum fractures has not been clarified in comparison of JP and 3DP constructs. This study analyzed the difference between the biomechanical stability of JP compared to 3DP and the role of an IACS in a standardized acetabular fracture model in a single-leg stance loading configuration.!##!Methods!#!In an artificial bone substitute pelvis model (Synbone!##!Results!#!In the pure placement, 3DP provided significantly superior stability when compared to JP. Augmentation of JP by IACS increased the stability significantly, up to the level of 3DP alone, whereas augmentation of the 3DP did not result in further increase of overall stability.!##!Conclusion!#!The anatomically shaped plate alone provides a superior biomechanical stability in fixation of an anterior column and posterior hemi-transverse fracture model. In a JP fixation the augmentation by IACS provides similar strength as the anatomically shaped 3DP. By use of the anatomically shaped 3DP the need of a clinically risky application of IACS might be avoidable.!##!Level of evidence!#!IV, Experimental study.
1000 Sacherschließung
lokal Hip Fractures [MeSH]
lokal Acetabulum/surgery [MeSH]
lokal Humans [MeSH]
lokal Suprapectineal
lokal Biomechanical Phenomena [MeSH]
lokal Biomechanics
lokal Bone Substitutes [MeSH]
lokal Acetabulum/injuries [MeSH]
lokal Bone Screws [MeSH]
lokal Bone Plates [MeSH]
lokal Original Article
lokal Infra-acetabular screw
lokal Anatomically shaped plate
lokal Acetabular fracture
lokal Fracture Fixation, Internal/methods [MeSH]
lokal Fractures, Bone/surgery [MeSH]
lokal Spinal Fractures [MeSH]
lokal Fractures, Bone/diagnostic imaging [MeSH]
1000 Liste der Beteiligten
  1. https://frl.publisso.de/adhoc/uri/R3JhdWwsIEku|https://frl.publisso.de/adhoc/uri/TWFyaW50c2NoZXYsIEku|https://frl.publisso.de/adhoc/uri/UGl6YW5pcywgQS4=|https://frl.publisso.de/adhoc/uri/SGVyYXRoLCBTLiBDLg==|https://frl.publisso.de/adhoc/uri/UG9obGVtYW5uLCBULg==|https://orcid.org/0000-0003-2007-0404
1000 Hinweis
  • DeepGreen-ID: ac2e441f1a764b06842a5ced46278d9d ; metadata provieded by: DeepGreen (https://www.oa-deepgreen.de/api/v1/), LIVIVO search scope life sciences (http://z3950.zbmed.de:6210/livivo), Crossref Unified Resource API (https://api.crossref.org/swagger-ui/index.html), to.science.api (https://frl.publisso.de/), ZDB JSON-API (beta) (https://zeitschriftendatenbank.de/api/), lobid - Dateninfrastruktur für Bibliotheken (https://lobid.org/resources/search)
1000 Label
1000 Dateien
1000 Objektart article
1000 Beschrieben durch
1000 @id frl:6448465.rdf
1000 Erstellt am 2023-05-03T18:32:08.141+0200
1000 Erstellt von 322
1000 beschreibt frl:6448465
1000 Zuletzt bearbeitet 2023-10-20T22:46:04.328+0200
1000 Objekt bearb. Fri Oct 20 22:46:04 CEST 2023
1000 Vgl. frl:6448465
1000 Oai Id
  1. oai:frl.publisso.de:frl:6448465 |
1000 Sichtbarkeit Metadaten public
1000 Sichtbarkeit Daten public
1000 Gegenstand von

View source