Download
s00586-021-06822-3.pdf 1,58MB
WeightNameValue
1000 Titel
  • Surgical nuances and construct patterns influence construct stiffness in C1-2 stabilizations: a biomechanical study of C1-2 gapping and advanced C1-2 fixation
1000 Autor/in
  1. Koller, Heiko |
  2. Hartmann, Sebastian |
  3. Raphael, Gmeiner |
  4. Schmölz, Werner |
  5. Orban, Christoph |
  6. Thome, Claudius |
1000 Erscheinungsjahr 2021
1000 Publikationstyp
  1. Artikel |
1000 Online veröffentlicht
  • 2021-04-23
1000 Erschienen in
1000 Quellenangabe
  • 30(6):1596-1606
1000 Copyrightjahr
  • 2021
1000 Lizenz
1000 Verlagsversion
  • https://doi.org/10.1007/s00586-021-06822-3 |
1000 Publikationsstatus
1000 Sprache der Publikation
1000 Abstract/Summary
  • Purpose!#!Stabilization of C1-2 using a Harms-Goel construct with 3.5 mm titanium (Ti) rods has been established as a standard of reference (SOR). A reduction in craniocervical deformities can indicate increased construct stiffness at C1-2. A reduction in C1-2 can result in C1-2 joint gapping. Therefore, the authors sought to study the biomechanical consequences of C1-2 gapping on construct stiffness using different instrumentations, including a novel 6-screw/3-rod (6S3R) construct, to compare the results to the SOR. We hypothesized that different instrument pattern will reveal significant differences in reduction in ROM among constructs tested.!##!Methods!#!The range of motion (ROM) of instrumented C1-2 polyamide models was analyzed in a six-degree-of-freedom spine tester. The models were loaded with pure moments (2.0 Nm) in axial rotation (AR), flexion extension (FE), and lateral bending (LB). Comparisons of C1-2 construct stiffness among the constructs included variations in rod diameter (3.5 mm vs. 4.0 mm), rod material (Ti. vs. CoCr) and a cross-link (CLX). Construct stiffness was tested with C1-2 facets in contact (Contact Group) and in a 2 mm distracted position (Gapping Group). The ROM (°) was recorded and reported as a percentage of ROM (%ROM) normalized to the SOR. A difference > 30% between the SOR and the %ROM among the constructs was defined as significant.!##!Results!#!Among all constructs, an increase in construct stiffness up to 50% was achieved with the addition of CLX, particularly with a 6S3R construct. These differences showed the greatest effect for the CLX in AR testing and for the 6S3R construct in FE and AR testing. Among all constructs, C1-2 gapping resulted in a significant loss of construct stiffness. A protective effect was shown for the CLX, particularly using a 6S3R construct in AR and FE testing. The selection of rod diameter (3.5 mm vs. 4.0 mm) and rod material (Ti vs. CoCr) did show a constant trend but did not yield significance.!##!Conclusion!#!This study is the first to show the loss of construct stiffness at C1-2 with gapping and increased restoration of stability using CLX and 6S3R constructs. In the correction of a craniocervical deformity, nuances in the surgical technique and advanced instrumentation may positively impact construct stability.
1000 Sacherschließung
lokal Construct stability
lokal Biomechanical study
lokal Humans [MeSH]
lokal Biomechanical Phenomena [MeSH]
lokal Bone Screws [MeSH]
lokal Spinal Fusion [MeSH]
lokal ROM
lokal Original Article
lokal Mechanical testing
lokal Implant pattern
lokal Range of Motion, Articular [MeSH]
lokal Cadaver [MeSH]
lokal C1-2
lokal Gapping
lokal Instrumentation
1000 Liste der Beteiligten
  1. https://frl.publisso.de/adhoc/uri/S29sbGVyLCBIZWlrbw==|https://frl.publisso.de/adhoc/uri/SGFydG1hbm4sIFNlYmFzdGlhbg==|https://frl.publisso.de/adhoc/uri/UmFwaGFlbCwgR21laW5lcg==|https://frl.publisso.de/adhoc/uri/U2NobcO2bHosIFdlcm5lcg==|https://frl.publisso.de/adhoc/uri/T3JiYW4sIENocmlzdG9waA==|https://frl.publisso.de/adhoc/uri/VGhvbWUsIENsYXVkaXVz
1000 Hinweis
  • DeepGreen-ID: 6a592634d71345079bb96da9c1828407 ; metadata provieded by: DeepGreen (https://www.oa-deepgreen.de/api/v1/), LIVIVO search scope life sciences (http://z3950.zbmed.de:6210/livivo), Crossref Unified Resource API (https://api.crossref.org/swagger-ui/index.html), to.science.api (https://frl.publisso.de/), ZDB JSON-API (beta) (https://zeitschriftendatenbank.de/api/), lobid - Dateninfrastruktur für Bibliotheken (https://lobid.org/resources/search)
1000 Label
1000 Dateien
1000 Objektart article
1000 Beschrieben durch
1000 @id frl:6449544.rdf
1000 Erstellt am 2023-05-09T09:19:38.138+0200
1000 Erstellt von 322
1000 beschreibt frl:6449544
1000 Zuletzt bearbeitet Sat Oct 21 01:06:38 CEST 2023
1000 Objekt bearb. Sat Oct 21 01:06:38 CEST 2023
1000 Vgl. frl:6449544
1000 Oai Id
  1. oai:frl.publisso.de:frl:6449544 |
1000 Sichtbarkeit Metadaten public
1000 Sichtbarkeit Daten public
1000 Gegenstand von