Download
s00405-021-06737-1.pdf 684,50KB
WeightNameValue
1000 Titel
  • UV light-based reprocessing of flexible endoscopes without working channel in Oto-Rhino-Laryngology: an effective method?
1000 Autor/in
  1. Rudhart, Stefan |
  2. Günther, Frank |
  3. Dapper, Laura |
  4. Thangavelu, Kruthika |
  5. Geisthoff, Urban W. |
  6. Stankovic, Petar |
  7. Wilhelm, Thomas |
  8. Stuck, Boris A. |
  9. Hoch, Stephan |
1000 Erscheinungsjahr 2021
1000 Publikationstyp
  1. Artikel |
1000 Online veröffentlicht
  • 2021-03-13
1000 Erschienen in
1000 Quellenangabe
  • 278(10):4075-4080
1000 Copyrightjahr
  • 2021
1000 Lizenz
1000 Verlagsversion
  • https://doi.org/10.1007/s00405-021-06737-1 |
  • https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8382609/ |
1000 Publikationsstatus
1000 Sprache der Publikation
1000 Abstract/Summary
  • Background!#!Reprocessing of flexible endoscopes (FEs) is often expensive, time consuming, and becomes increasingly complex, due to rising demands of hygiene. After beneficial results in reprocessing of rigid endoscopes using Impelux™ UV-C light technology, we tested the same method for reprocessing of FEs without working channel.!##!Materials and methods!#!Testing was performed on FEs without working channel after routine clinical use (transnasal flexible endoscopy). Disinfection consisted of mechanical precleaning and 60 s exposure to Impelux™ UV-C light technology. Bacterial contamination was tested on 50 FEs before and after disinfection. Further 50 FEs regarding protein residuals. The absolute effectiveness of the D60 was tested on 50 test bodies (RAMS) with a standardized contamination of 10!##!Results!#!The FEs were contaminated with a high average value of 916.7 CFU (± 1057 CFU) after clinical usage. After reprocessing, an average contamination of 2.8 CFU (± 1.6) on 14% (n = 7) of the FEs was detected consisting of non-pathogenic species, the remaining FE were sterile. After reprocessing, all FEs were protein-free (< 1 μg). The artificially contaminated test bodies showed no remaining bacterial contamination after disinfection, resulting in an average absolute germ reduction of about 10!##!Conclusion!#!Impelux™ UV-C light technology efficiently reduces bacterial contamination of FEs and might be useful in daily practice.
1000 Sacherschließung
lokal Endoscopes [MeSH]
lokal Disinfection [MeSH]
lokal Disinfection
lokal Animals [MeSH]
lokal Sheep [MeSH]
lokal Miscellaneous
lokal Male [MeSH]
lokal Sterilization
lokal Otolaryngology [MeSH]
lokal Ultraviolet Rays [MeSH]
lokal Equipment Contamination/prevention
lokal Microbiology
lokal Hygiene
1000 Liste der Beteiligten
  1. https://orcid.org/0000-0003-3580-4255|https://frl.publisso.de/adhoc/uri/R8O8bnRoZXIsIEZyYW5r|https://frl.publisso.de/adhoc/uri/RGFwcGVyLCBMYXVyYQ==|https://frl.publisso.de/adhoc/uri/VGhhbmdhdmVsdSwgS3J1dGhpa2E=|https://frl.publisso.de/adhoc/uri/R2Vpc3Rob2ZmLCBVcmJhbiBXLg==|https://frl.publisso.de/adhoc/uri/U3RhbmtvdmljLCBQZXRhcg==|https://frl.publisso.de/adhoc/uri/V2lsaGVsbSwgVGhvbWFz|https://frl.publisso.de/adhoc/uri/U3R1Y2ssIEJvcmlzIEEu|https://frl.publisso.de/adhoc/uri/SG9jaCwgU3RlcGhhbg==
1000 Hinweis
  • DeepGreen-ID: d993327077c143d0b3fd2341a03f62ca ; metadata provieded by: DeepGreen (https://www.oa-deepgreen.de/api/v1/), LIVIVO search scope life sciences (http://z3950.zbmed.de:6210/livivo), Crossref Unified Resource API (https://api.crossref.org/swagger-ui/index.html), to.science.api (https://frl.publisso.de/), ZDB JSON-API (beta) (https://zeitschriftendatenbank.de/api/), lobid - Dateninfrastruktur für Bibliotheken (https://lobid.org/resources/search)
1000 Label
1000 Dateien
1000 Objektart article
1000 Beschrieben durch
1000 @id frl:6450448.rdf
1000 Erstellt am 2023-05-09T12:24:54.584+0200
1000 Erstellt von 322
1000 beschreibt frl:6450448
1000 Zuletzt bearbeitet 2023-10-21T03:31:36.557+0200
1000 Objekt bearb. Sat Oct 21 03:31:36 CEST 2023
1000 Vgl. frl:6450448
1000 Oai Id
  1. oai:frl.publisso.de:frl:6450448 |
1000 Sichtbarkeit Metadaten public
1000 Sichtbarkeit Daten public
1000 Gegenstand von

View source