Download
am10051e2019.pdf 360,78KB
WeightNameValue
1000 Titel
  • Benzyl alcohol – Method for the determination of benzyl alcohol in workplace air using gas chromatography after thermal desorption : Air Monitoring Methods, 2018
1000 Titelzusatz
  • Air Monitoring Methods – Benzyl alcohol – Method for the determination of benzyl alcohol in workplace air using gas chromatography after thermal desorption
1000 Beteiligung
Deutsche Forschungsgemeinschaft. Ständige Senatskommission zur Prüfung Gesundheitsschädlicher Arbeitsstoffe (herausgebende Körperschaft) |
1000 Autor/in
  1. Deutsche Forschungsgemeinschaft. Ständige Senatskommission zur Prüfung Gesundheitsschädlicher Arbeitsstoffe |
  2. Tschickardt, Michael |
  3. Krämer, Wilhelm |
  4. Schmitt, Ralf |
  5. Hebisch, Ralph |
  6. Brock, Thomas Helmut |
  7. Hartwig, Andrea |
  8. MAK Commission |
1000 Erscheinungsjahr 2019
1000 Publikationstyp
  1. Artikel |
1000 Online veröffentlicht
  • 2019-07-25
1000 Erschienen in
1000 Quellenangabe
  • 4(3):1667-1682
1000 FRL-Sammlung
1000 Copyrightjahr
  • 2019
1000 Lizenz
1000 Verlagsversion
  • https://doi.org/10.1002/3527600418.am10051e2019 |
1000 Publikationsstatus
1000 Begutachtungsstatus
1000 Sprache der Publikation
1000 Abstract/Summary
  • This analytical method is a validated measurement procedure for the determination of benzyl alcohol [100‐51‐6] in workplace air in a concentration range of one tenth up to twice the currently valid MAK value of 22 mg/m3. With this method simultaneously airborne benzyl alcohol in the gaseous state as well as particles are collected. Sampling is performed by drawing a defined volume of air through a quartz fibre filter located in the sampling head with an adsorption tube filled with Tenax TA® connected downstream using a suitable pump. The flow rate is set to 0.066 L/min with a recommended air sample volume of approx. 4 litres. The collected benzyl alcohol is thermally desorbed and then analysed by means of gas chromatography using two detectors, a FID for quantification and a MSD to recognise potential interferences. The quantitative determination is based on a calibration function obtained by means of a multi‐point calibration. The limit of quantification is 0.24 mg/m3 based on an air sample of approx. 4 litres.
1000 Sacherschließung
lokal workplace measurement
lokal Luftanalysen-Methoden
lokal GGP-Mini
lokal hazardous substances
lokal workplace monitoring
lokal thermal desorption
lokal benzyl alcohol
lokal GC-MSD
lokal gas chromatography
lokal 100-51-6
lokal GC-FID
lokal Tenax TA
lokal Air Monitoring Methods
lokal air analysis
lokal glass fibre filter
1000 Fächerklassifikation (DDC)
1000 DOI 10.4126/FRL01-006453709 |
1000 Liste der Beteiligten
  1. VHNjaGlja2FyZHQsIE1pY2hhZWw=|S3LDpG1lciwgV2lsaGVsbQ==|U2NobWl0dCwgUmFsZg==|SGViaXNjaCwgUmFscGg=|QnJvY2ssIFRob21hcyBIZWxtdXQ=|SGFydHdpZywgQW5kcmVh|TUFLIENvbW1pc3Npb24=
1000 Label
1000 Dateien
1000 Objektart article
1000 Beschrieben durch
1000 @id frl:6453709.rdf
1000 Erstellt am 2023-08-22T09:41:05.013+0200
1000 Erstellt von 335
1000 beschreibt frl:6453709
1000 Zuletzt bearbeitet 2024-01-04T12:09:47.808+0100
1000 Objekt bearb. Thu Jan 04 12:09:47 CET 2024
1000 Vgl. frl:6453709
1000 Oai Id
  1. oai:frl.publisso.de:frl:6453709 |
1000 Sichtbarkeit Metadaten public
1000 Sichtbarkeit Daten public
1000 Gegenstand von

View source