  • Evaluation of an occipito-cervico fusion with a new implant design: a biomechanical study
1000 Autor/in
  1. Migliorini, Filippo |
  2. Baroncini, Alice |
  3. El Mansy, Yasser |
  4. Quack, Valentin |
  5. Prescher, Andreas |
  6. Mischer, Max |
  7. Greven, Johannes |
  8. Tingart, Markus |
  9. Eschweiler, Jörg |
1000 Erscheinungsjahr 2021
1000 Publikationstyp
  1. Artikel |
1000 Online veröffentlicht
  • 2021-03-06
1000 Erschienen in
1000 Quellenangabe
  • 22(1):250
1000 Copyrightjahr
  • 2021
1000 Lizenz
1000 Verlagsversion
  • https://doi.org/10.1186/s12891-021-04112-z |
  • https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7937312/ |
1000 Publikationsstatus
1000 Sprache der Publikation
1000 Abstract/Summary
  • Background!#!A novel implant for occipitocervical fusion consisting of a median plate with an additional hook inserting in the foramen magnum was tested. Aim of this study was to test the stability of a new implant for occipitocervical fusion against the already available and employed median plate implant without hook.!##!Material and method!#!36 rigid polyurethane foams occipital artificial bones were used. The two occipital implants, namely the occipital plate with hook (Group 1) and the one without hook (Group 2), were applied to the artificial occiput trough three occipital screws and ensured into the experimental setup trough a crossbar. The test parameters were set using the testing machine software as follows: (1) test speed: 10 mm/ min, with 25 mm/ min maximum; (2) preload: 5 N; (3) force switch-off threshold: 90% force drop from F_max. Failure force and path were recorded. Failure force is defined as the maximum reaction force under which failure occurs (F_max), while failure path is the travel path during which failure occurs (dL).!##!Results!#!Group 1 (plate with hook) showed a mean failure force of 459.3 ± 35.9 N and a mean failure path of 5.8 ± 0.3 mm Group 2 (plate without hook) showed a mean failure force of 323.9 ± 20.2 N and a mean failure path of 7.2 ± 0.4 mm. The Shapiro-Wilk test score was not significant (P >  0.1), assuming that data were normally distributed. Group 1 had a statistically significant greater F_max (+ 135.37; P >  0.0001) and less dL (- 1.52; P > 0.0001) compared to group 2.!##!Conclusions!#!Medial plates with foramen magnum hooks showed to be more stable that plates without a hook. These new implants may represent a new tool in OCJ fixation, but further studies are required to investigate their behavior in an anatomical setting.
1000 Sacherschließung
lokal Rheumatology
lokal Occipital Bone/diagnostic imaging [MeSH]
lokal Humans [MeSH]
lokal Biomechanical Phenomena [MeSH]
lokal Sports Medicine
lokal Bone Screws [MeSH]
lokal Orthopedics
lokal Spinal Fusion [MeSH]
lokal Bone Plates [MeSH]
lokal Epidemiology
lokal Rehabilitation
lokal Occipital Bone/surgery [MeSH]
lokal Internal Medicine
lokal Research Article
1000 Liste der Beteiligten
  1. https://orcid.org/0000-0001-7220-1221|https://frl.publisso.de/adhoc/uri/QmFyb25jaW5pLCBBbGljZQ==|https://frl.publisso.de/adhoc/uri/RWwgTWFuc3ksIFlhc3Nlcg==|https://frl.publisso.de/adhoc/uri/UXVhY2ssIFZhbGVudGlu|https://frl.publisso.de/adhoc/uri/UHJlc2NoZXIsIEFuZHJlYXM=|https://frl.publisso.de/adhoc/uri/TWlzY2hlciwgTWF4|https://frl.publisso.de/adhoc/uri/R3JldmVuLCBKb2hhbm5lcw==|https://frl.publisso.de/adhoc/uri/VGluZ2FydCwgTWFya3Vz|https://frl.publisso.de/adhoc/uri/RXNjaHdlaWxlciwgSsO2cmc=
1000 Hinweis
  • DeepGreen-ID: 9b90c16760fd4a3796e6e0aa16608f95 ; metadata provieded by: DeepGreen (https://www.oa-deepgreen.de/api/v1/), LIVIVO search scope life sciences (http://z3950.zbmed.de:6210/livivo), Crossref Unified Resource API (https://api.crossref.org/swagger-ui/index.html), to.science.api (https://frl.publisso.de/), ZDB JSON-API (beta) (https://zeitschriftendatenbank.de/api/), lobid - Dateninfrastruktur für Bibliotheken (https://lobid.org/resources/search)
1000 Label
1000 Dateien
1000 Objektart article
1000 Beschrieben durch
1000 @id frl:6463785.rdf
1000 Erstellt am 2023-11-15T21:16:54.728+0100
1000 Erstellt von 322
1000 beschreibt frl:6463785
1000 Zuletzt bearbeitet 2023-11-30T22:20:11.121+0100
1000 Objekt bearb. Thu Nov 30 22:20:11 CET 2023
1000 Vgl. frl:6463785
1000 Oai Id
  1. oai:frl.publisso.de:frl:6463785 |
1000 Sichtbarkeit Metadaten public
1000 Sichtbarkeit Daten public
1000 Gegenstand von

View source