Download
s00392-020-01767-x.pdf 1,14MB
WeightNameValue
1000 Titel
  • Pitfalls of the S-ICD therapy: experiences from a large tertiary centre
1000 Autor/in
  1. Willy, Kevin |
  2. Reinke, Florian |
  3. Rath, Benjamin |
  4. Ellermann, Christian |
  5. Wolfes, Julian |
  6. Bögeholz, Nils |
  7. Köbe, Julia |
  8. Eckardt, Lars |
  9. Frommeyer, Gerrit |
1000 Erscheinungsjahr 2020
1000 Publikationstyp
  1. Artikel |
1000 Online veröffentlicht
  • 2020-11-01
1000 Erschienen in
1000 Quellenangabe
  • 110(6):861-867
1000 Copyrightjahr
  • 2020
1000 Lizenz
1000 Verlagsversion
  • https://doi.org/10.1007/s00392-020-01767-x |
  • https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8166696/ |
1000 Publikationsstatus
1000 Sprache der Publikation
1000 Abstract/Summary
  • Aim!#!The subcutaneous ICD (S-ICD) has evolved to a potential first option for many patients who have to be protected from sudden cardiac death. Many trials have underlined a similar performance regarding its effectiveness in relation to transvenous ICDs and have shown the expected benefits concerning infective endocarditis and lead failure. However, there have also been problems due to the peculiarities of the device, such as oversensing and myopotentials. In this study, we present patients from a large tertiary centre suffering from complications with an S-ICD and propose possible solutions.!##!Methods and results!#!All S-ICD patients who experienced complications related to the device (n = 40) of our large-scale single-centre S-ICD registry (n = 351 patients) were included in this study. Baseline characteristics, complications occurring and solutions to these problems were documented over a mean follow-up of 50 months. In most cases (n = 23), patients suffered from oversensing (18 cases with T wave or P wave oversensing, 5 due to myopotentials). Re-programming successfully prevented further oversensing episode in 13/23 patients. In 9 patients, generator or lead-related complications, mostly due to infectious reasons (5/9), occurred. Further problems consisted of ineffective shocks in one patient and need for antibradycardia stimulation in 2 patients and indication for CRT in 2 other patients. In total, the S-ICD had to be extracted in 10 patients. 7 of them received a tv-ICD subsequently, 3 patients refused re-implantation of any ICD. One other patient kept the ICD but had antitachycardic therapy deactivated due to inappropriate shocks for myopotential oversensing.!##!Conclusion!#!The S-ICD is a valuable option for many patients for the prevention of sudden cardiac death. Nonetheless, certain problems are immanent to the S-ICD (limited re-programming options, size of the generator) and should be addressed in future generations of the S-ICD.
1000 Sacherschließung
lokal S-ICD
lokal Electrodes, Implanted/statistics
lokal Female [MeSH]
lokal Pitfalls
lokal Defibrillators, Implantable [MeSH]
lokal ICD
lokal Adult [MeSH]
lokal Humans [MeSH]
lokal Practice Guidelines as Topic [MeSH]
lokal Retrospective Studies [MeSH]
lokal Ventricular arrhythmia
lokal Lead dysfunction
lokal Subcutaneous ICD
lokal Male [MeSH]
lokal Troubleshooting
lokal Death, Sudden, Cardiac/prevention
lokal Original Paper
lokal Tertiary Care Centers/statistics
1000 Liste der Beteiligten
  1. https://orcid.org/0000-0003-1259-3213|https://frl.publisso.de/adhoc/uri/UmVpbmtlLCBGbG9yaWFu|https://frl.publisso.de/adhoc/uri/UmF0aCwgQmVuamFtaW4=|https://frl.publisso.de/adhoc/uri/RWxsZXJtYW5uLCBDaHJpc3RpYW4=|https://frl.publisso.de/adhoc/uri/V29sZmVzLCBKdWxpYW4=|https://frl.publisso.de/adhoc/uri/QsO2Z2Vob2x6LCBOaWxz|https://frl.publisso.de/adhoc/uri/S8O2YmUsIEp1bGlh|https://frl.publisso.de/adhoc/uri/RWNrYXJkdCwgTGFycw==|https://frl.publisso.de/adhoc/uri/RnJvbW1leWVyLCBHZXJyaXQ=
1000 Hinweis
  • DeepGreen-ID: 71bf59cfb8f64456b433c751259036fd ; metadata provieded by: DeepGreen (https://www.oa-deepgreen.de/api/v1/), LIVIVO search scope life sciences (http://z3950.zbmed.de:6210/livivo), Crossref Unified Resource API (https://api.crossref.org/swagger-ui/index.html), to.science.api (https://frl.publisso.de/), ZDB JSON-API (beta) (https://zeitschriftendatenbank.de/api/), lobid - Dateninfrastruktur für Bibliotheken (https://lobid.org/resources/search)
1000 Label
1000 Dateien
1000 Objektart article
1000 Beschrieben durch
1000 @id frl:6469994.rdf
1000 Erstellt am 2023-11-18T03:10:40.437+0100
1000 Erstellt von 322
1000 beschreibt frl:6469994
1000 Zuletzt bearbeitet 2023-12-01T10:57:14.498+0100
1000 Objekt bearb. Fri Dec 01 10:57:14 CET 2023
1000 Vgl. frl:6469994
1000 Oai Id
  1. oai:frl.publisso.de:frl:6469994 |
1000 Sichtbarkeit Metadaten public
1000 Sichtbarkeit Daten public
1000 Gegenstand von

View source