Download
00405_2023_Article_8212.pdf 553,09KB
WeightNameValue
1000 Titel
  • Evaluation of insertion quality of a slim perimodiolar electrode array
1000 Autor/in
  1. Beck, R. |
  2. Aschendorff, A. |
  3. Arndt, S. |
  4. Hildenbrand, T. |
  5. Rauch, A. K. |
  6. Ketterer, Manuel Christoph |
1000 Verlag Springer Berlin Heidelberg
1000 Erscheinungsjahr 2023
1000 Publikationstyp
  1. Artikel |
1000 Online veröffentlicht
  • 2023-09-29
1000 Erschienen in
1000 Quellenangabe
  • 281(3):1215-1220
1000 Copyrightjahr
  • 2023
1000 Lizenz
1000 Verlagsversion
  • https://doi.org/10.1007/s00405-023-08212-5 |
  • https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10858093/ |
1000 Publikationsstatus
1000 Begutachtungsstatus
1000 Sprache der Publikation
1000 Abstract/Summary
  • <jats:title>Abstract</jats:title><jats:sec> <jats:title>Objectives</jats:title> <jats:p>The influence of cochlear morphology and electrode array design on scalar position and dislocation rates is of great interest in CI surgery. The aim of this study is to evaluate scalar position and specific points of dislocation in relation to cochlear morphology in patients implanted with a new slim perimodiolar electrode array.</jats:p> </jats:sec><jats:sec> <jats:title>Materials and methods</jats:title> <jats:p>Patients were implanted using the slim modiolar electrode array (= SMA) (= 532/632 electrode array of Cochlear<jats:sup>™</jats:sup>). Postoperative imaging was performed via cone beam computed tomography (CBCT) and the scans were analyzed regarding cochlear morphology (distances A and B and cochlear height), scalar location of the electrode array, basal insertion depth and apical insertion angle. Furthermore, electrode array design and surgical protocols were evaluated.</jats:p> </jats:sec><jats:sec> <jats:title>Results</jats:title> <jats:p>81 ears implanted with the SMA were retrospectively included. We evaluated 3 electrode array tip fold over intraoperatively via X-ray imaging and performed revision during the same surgery. The CBCT scans showed 76 initial scala tympani (ST) insertions without dislocation. Two ears showed a dislocated array, one at 77° and the other at 163°. Three arrays were inserted into scala vestibuli (SV) via cochleostomy. These patients showed no signs of obliteration. Cochlear morphology showed no influence on angular insertion depth and scalar position.</jats:p> </jats:sec><jats:sec> <jats:title>Conclusions</jats:title> <jats:p>The SMA showed a very low rate of scalar dislocations due to its slim electrode array design (2.7%). We could find a learning curve regarding the handling and the risk of dislocation and tip fold over with this electrode array. The rate of intraoperative tip fold over detection via X-ray imaging was 3.7%. Therefore, we highly recommend X-ray imaging and transimpedance matrix measurements within the surgery protocol. Scala vestibuli insertions happened in patients with cochleostomy only. We could identify two specific points of dislocation depending on electrode array design.</jats:p> </jats:sec>
1000 Sacherschließung
lokal Humans [MeSH]
lokal Electrodes, Implanted [MeSH]
lokal Perimodiolar array
lokal Cochlear implant surgery
lokal Retrospective Studies [MeSH]
lokal Cochlea/surgery [MeSH]
lokal Otology
lokal Scala Tympani/surgery [MeSH]
lokal Cochlear Implantation/methods [MeSH]
lokal Electrode array design
lokal Scalar position
lokal Cochlear Implants [MeSH]
lokal Cochlea/anatomy
lokal Cochlea/diagnostic imaging [MeSH]
1000 Fächerklassifikation (DDC)
1000 Liste der Beteiligten
  1. https://frl.publisso.de/adhoc/uri/QmVjaywgUi4=|https://frl.publisso.de/adhoc/uri/QXNjaGVuZG9yZmYsIEEu|https://frl.publisso.de/adhoc/uri/QXJuZHQsIFMu|https://frl.publisso.de/adhoc/uri/SGlsZGVuYnJhbmQsIFQu|https://frl.publisso.de/adhoc/uri/UmF1Y2gsIEEuIEsu|https://orcid.org/0000-0003-0391-6607
1000 Hinweis
  • DeepGreen-ID: 2cfbe36f0a214e238575872cfe8bacba ; metadata provieded by: DeepGreen (https://www.oa-deepgreen.de/api/v1/), LIVIVO search scope life sciences (http://z3950.zbmed.de:6210/livivo), Crossref Unified Resource API (https://api.crossref.org/swagger-ui/index.html), to.science.api (https://frl.publisso.de/), ZDB JSON-API (beta) (https://zeitschriftendatenbank.de/api/), lobid - Dateninfrastruktur für Bibliotheken (https://lobid.org/resources/search)
1000 Label
1000 Förderer
  1. Universitätsklinikum Freiburg |
1000 Fördernummer
  1. -
1000 Förderprogramm
  1. -
1000 Dateien
  1. Evaluation of insertion quality of a slim perimodiolar electrode array
1000 Förderung
  1. 1000 joinedFunding-child
    1000 Förderer Universitätsklinikum Freiburg |
    1000 Förderprogramm -
    1000 Fördernummer -
1000 Objektart article
1000 Beschrieben durch
1000 @id frl:6518226.rdf
1000 Erstellt am 2025-07-05T09:45:40.445+0200
1000 Erstellt von 322
1000 beschreibt frl:6518226
1000 Zuletzt bearbeitet 2025-08-19T12:32:07.413+0200
1000 Objekt bearb. Tue Aug 19 12:32:07 CEST 2025
1000 Vgl. frl:6518226
1000 Oai Id
  1. oai:frl.publisso.de:frl:6518226 |
1000 Sichtbarkeit Metadaten public
1000 Sichtbarkeit Daten public
1000 Gegenstand von

View source