Download
s00270-020-02584-5.pdf 366,33KB
WeightNameValue
1000 Titel
  • MR-Guided High-Power Microwave Ablation in Hepatic Malignancies: Initial Results in Clinical Routine
1000 Autor/in
  1. Winkelmann, Moritz T. |
  2. Gohla, Georg |
  3. Kübler, Jens |
  4. Weiß, Jakob |
  5. Clasen, Stephan |
  6. Nikolaou, Konstantin |
  7. Hoffmann, Rüdiger |
1000 Erscheinungsjahr 2020
1000 Publikationstyp
  1. Artikel |
1000 Online veröffentlicht
  • 2020-07-22
1000 Erschienen in
1000 Quellenangabe
  • 43(11):1631-1638
1000 Copyrightjahr
  • 2020
1000 Lizenz
1000 Verlagsversion
  • https://doi.org/10.1007/s00270-020-02584-5 |
  • https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7591402/ |
1000 Publikationsstatus
1000 Begutachtungsstatus
1000 Sprache der Publikation
1000 Abstract/Summary
  • Purpose!#!Evaluation of technique effectiveness, patient safety and ablation parameters of MR-guided microwave ablation in hepatic malignancies using an MR-conditional high-power microwave ablation system.!##!Materials and methods!#!Institutional review board approval and informed patient consent were obtained. Patients who underwent MR-guided microwave ablation of hepatic malignancies in a 1.5T wide-bore scanner using a perfusion-cooled high-power microwave ablation system with a maximum generator power of 150 W were included. Ablation parameters comprising procedure durations, net ablation duration, applicator positions and ablation zone dimensions were recorded. Adverse events were classified according to the CIRSE classification system. Technique effectiveness was assessed after 1 month. Follow-up was conducted with contrast-enhanced MRI and ranged from 1 to 20 months (mean: 6.1 ± 5.4 months).!##!Results!#!Twenty-one consecutive patients (age: 63.4 ± 10.5 years; 5 female) underwent 22 procedures for 28 tumours (9 hepatocellular carcinomas, 19 metastases) with a mean tumour diameter of 14.6 ± 5.4 mm (range: 6-24 mm). Technique effectiveness was achieved in all lesions. Tumours were treated using 1.7 ± 0.7 applicator positions (range: 1-3). Mean energy and ablation duration per tumour were 75.3 ± 35.4 kJ and 13.3 ± 6.2 min, respectively. Coagulation zone short- and long-axis diameters were 29.1 ± 6.4 mm and 39.9 ± 7.4 mm, respectively. Average procedure duration was 146.4 ± 26.2 min (range: 98-187 min). One minor complication was reported. Five patients developed new tumour manifestations in the untreated liver. Local tumour progression was not observed during initial follow-up.!##!Conclusion!#!MR-guided high-power microwave ablation provides safe and effective treatment of hepatic malignancies with short ablation times and within acceptable procedure durations.
1000 Sacherschließung
lokal Female [MeSH]
lokal Clinical Investigation
lokal Therapy, Computer-Assisted/methods [MeSH]
lokal Aged [MeSH]
lokal Liver Neoplasms/diagnosis [MeSH]
lokal Adult [MeSH]
lokal Humans [MeSH]
lokal Liver Neoplasms/therapy [MeSH]
lokal Liver neoplasms
lokal Treatment Outcome [MeSH]
lokal Carcinoma, Hepatocellular/diagnosis [MeSH]
lokal Duration of therapy
lokal Middle Aged [MeSH]
lokal Carcinoma, Hepatocellular/therapy [MeSH]
lokal Neoplasm Staging [MeSH]
lokal Ablation Techniques/methods [MeSH]
lokal Male [MeSH]
lokal Patient Safety [MeSH]
lokal Microwaves/therapeutic use [MeSH]
lokal Ablation techniques
lokal Magnetic Resonance Imaging/methods [MeSH]
lokal Microwave ablation
lokal Interventional magnetic resonance imaging
lokal Non-Vascular Interventions
1000 Liste der Beteiligten
  1. https://frl.publisso.de/adhoc/uri/V2lua2VsbWFubiwgTW9yaXR6IFQu|https://frl.publisso.de/adhoc/uri/R29obGEsIEdlb3Jn|https://frl.publisso.de/adhoc/uri/S8O8YmxlciwgSmVucw==|https://frl.publisso.de/adhoc/uri/V2Vpw58sIEpha29i|https://frl.publisso.de/adhoc/uri/Q2xhc2VuLCBTdGVwaGFu|https://frl.publisso.de/adhoc/uri/Tmlrb2xhb3UsIEtvbnN0YW50aW4=|https://orcid.org/0000-0002-1383-4798
1000 Hinweis
  • DeepGreen-ID: 3587c452fbbc4e81a515343bf90969ec ; metadata provieded by: DeepGreen (https://www.oa-deepgreen.de/api/v1/), LIVIVO search scope life sciences (http://z3950.zbmed.de:6210/livivo), Crossref Unified Resource API (https://api.crossref.org/swagger-ui/index.html), to.science.api (https://frl.publisso.de/), ZDB JSON-API (beta) (https://zeitschriftendatenbank.de/api/), lobid - Dateninfrastruktur für Bibliotheken (https://lobid.org/resources/search)
1000 Label
1000 Dateien
1000 Objektart article
1000 Beschrieben durch
1000 @id frl:6470136.rdf
1000 Erstellt am 2023-11-18T04:08:10.441+0100
1000 Erstellt von 322