Download
41433_2024_Article_3261.pdf 2,17MB
WeightNameValue
1000 Titel
  • Anterior segment optical coherence tomography for the detection of silicone oil emulsification on the iris surface
1000 Autor/in
  1. Karsten, Mona |
  2. Morello, Matteo |
  3. Lau, Imke |
  4. Druchkiv, Vasyl |
  5. Lopes, Ines Valente |
  6. Dulz, Simon |
  7. Skevas, Christos |
  8. Spitzer, Martin Stephan |
  9. Mautone, Luca |
1000 Verlag Nature Publishing Group UK
1000 Erscheinungsjahr 2024
1000 Publikationstyp
  1. Artikel |
1000 Online veröffentlicht
  • 2024-07-29
1000 Erschienen in
1000 Quellenangabe
  • 38(16):3175-3179
1000 Copyrightjahr
  • 2024
1000 Lizenz
1000 Verlagsversion
  • https://doi.org/10.1038/s41433-024-03261-4 |
  • https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11543845/ |
1000 Publikationsstatus
1000 Begutachtungsstatus
1000 Sprache der Publikation
1000 Abstract/Summary
  • <jats:title>Abstract</jats:title><jats:sec> <jats:title>Background</jats:title> <jats:p>To investigate the presence of silicone oil (SO)-emulsification on the anterior iris surface with anterior segment optical coherence tomography (AS-OCT).</jats:p> </jats:sec><jats:sec> <jats:title>Methods</jats:title> <jats:p>In this single-center cross-sectional study, vitrectomized eyes with SO tamponade that underwent AS-OCT imaging and gonioscopy examination during the postoperative follow-up visits, were reviewed.</jats:p> </jats:sec><jats:sec> <jats:title>Results</jats:title> <jats:p>45 eyes of 42 consecutive patients were included. In 35.6% of the eyes (<jats:italic>n</jats:italic> = 16) emulsified SO droplets were detected in the anterior chamber (AC) angle by gonioscopy and in 55.6% (<jats:italic>n</jats:italic> = 25) on the anterior iris surface by AS-OCT imaging. The presence of SO emulsifications in the AC-angle correlated with the presence of SO emulsifications on the anterior iris surface (OR = 13.4, 95% CI [2.179–82.130]; <jats:italic>p</jats:italic> = 0.005). The accuracy of the AS-OCT predicting the presence of SO in the AC-angle was 71.0% and the sensitivity was 87.5%. No significant association between SO droplets in the AC and other clinical parameters including endotamponade-duration or type of silicone oil were found. The presence of emulsified SO droplets on the anterior iris surface detected by AS-OCT was significantly correlated to postoperative IOP rise (<jats:italic>p</jats:italic> = 0.027).</jats:p> </jats:sec><jats:sec> <jats:title>Conclusion</jats:title> <jats:p>AS-OCT is a suitable method for the detection of SO on the anterior iris surface. SO droplets on the iris surface correlate with elevated postoperative IOP and with the presence of SO in the AC detected by gonioscopy, therefore AS-OCT might be used as a screening method for the detection of SO migration into the AC.</jats:p> </jats:sec><jats:sec> <jats:title>Trial registration number</jats:title> <jats:p>Institutional Review Board of the Hamburg Medical Chamber (Ethik-Kommission der Ärztekammer Hamburg): 2023-300372-WF.</jats:p> </jats:sec>
1000 Sacherschließung
lokal Vitrectomy [MeSH]
lokal Aged [MeSH]
lokal Gonioscopy [MeSH]
lokal /692/700/139
lokal Retinal Detachment/surgery [MeSH]
lokal Intraocular Pressure/physiology [MeSH]
lokal Iris/diagnostic imaging [MeSH]
lokal Male [MeSH]
lokal Retinal Detachment/diagnosis [MeSH]
lokal Silicone Oils [MeSH]
lokal Iris/pathology [MeSH]
lokal Iris Diseases/diagnostic imaging [MeSH]
lokal Anterior Eye Segment/diagnostic imaging [MeSH]
lokal Female [MeSH]
lokal Anterior Eye Segment/pathology [MeSH]
lokal Adult [MeSH]
lokal Humans [MeSH]
lokal Iris Diseases/diagnosis [MeSH]
lokal Retrospective Studies [MeSH]
lokal Middle Aged [MeSH]
lokal /692/700/1421/1846
lokal Cross-Sectional Studies [MeSH]
lokal Article
lokal Emulsions [MeSH]
lokal Tomography, Optical Coherence/methods [MeSH]
lokal Endotamponade [MeSH]
lokal article
1000 Fächerklassifikation (DDC)
1000 Liste der Beteiligten
  1. https://orcid.org/0009-0006-0023-9793|https://frl.publisso.de/adhoc/uri/TW9yZWxsbywgTWF0dGVv|https://frl.publisso.de/adhoc/uri/TGF1LCBJbWtl|https://frl.publisso.de/adhoc/uri/RHJ1Y2hraXYsIFZhc3ls|https://frl.publisso.de/adhoc/uri/TG9wZXMsIEluZXMgVmFsZW50ZQ==|https://frl.publisso.de/adhoc/uri/RHVseiwgU2ltb24=|https://frl.publisso.de/adhoc/uri/U2tldmFzLCBDaHJpc3Rvcw==|https://frl.publisso.de/adhoc/uri/U3BpdHplciwgTWFydGluIFN0ZXBoYW4=|https://frl.publisso.de/adhoc/uri/TWF1dG9uZSwgTHVjYQ==
1000 Hinweis
  • DeepGreen-ID: 1e005453bd9d405b8289f945cd3231fd ; metadata provieded by: DeepGreen (https://www.oa-deepgreen.de/api/v1/), LIVIVO search scope life sciences (http://z3950.zbmed.de:6210/livivo), Crossref Unified Resource API (https://api.crossref.org/swagger-ui/index.html), to.science.api (https://frl.publisso.de/), ZDB JSON-API (beta) (https://zeitschriftendatenbank.de/api/), lobid - Dateninfrastruktur für Bibliotheken (https://lobid.org/resources/search)
1000 Label
1000 Dateien
1000 Objektart article
1000 Beschrieben durch
1000 @id frl:6517672.rdf
1000 Erstellt am 2025-07-05T05:57:03.730+0200
1000 Erstellt von 322
1000 beschreibt frl:6517672
1000 Zuletzt bearbeitet 2025-08-19T10:10:17.770+0200
1000 Objekt bearb. Tue Aug 19 10:10:17 CEST 2025
1000 Vgl. frl:6517672
1000 Oai Id
  1. oai:frl.publisso.de:frl:6517672 |
1000 Sichtbarkeit Metadaten public
1000 Sichtbarkeit Daten public
1000 Gegenstand von

View source